SIEMENS AXIOM Artis - AUTOMAP FUNCTION

FDA Recall Z-1108-04 · Posted 2004-07-20

Recall StatusTerminated
Product CodeIZI
Date Initiated2004-06-02
Date Terminated2004-10-21
Product Quantity127 units

Reason for Recall

Error appears in the AUTOMAP function, Module 2, when PERISTEEPING or PERIVISION is activated simultaneously that causes system to crash.

Firm Action

The recalling firm sent recall letter to their customers informing them of the software problem. The customers were informed that a Siemens Service Representative will visit each location to deactivate Module 2 of the AUTOMAP function per Update AX023/04S.

Root Cause

Other

Distribution Pattern

The recalled product has been distributed throughout the United States, such as: AK, AL, AZ, CA, DC, FL, GA, IA, IL, IN, KY, LA, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI,