UniCel Dxl Access Immunoassay Systems Reaction Vessels for Use with UniCel Dxl Access Immunoassay Systems, Part Number : 386167, Beckman Coulter, Inc, Brea CA 92822
FDA Recall Z-1108-2008 · Posted 2008-03-19
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeMMI
Date Initiated2007-07-19
Date Terminated2012-05-04
Product Quantity264 units in the US; 9 units in Canada
Reason for Recall
Erroneous Results - UniCel Dxl Access Immunoassay Systems Reaction Vessels have been found to be defective and there is a potential for an increase in signal leading to an erroneous result.
Firm Action
A Product Corrective Action (PCA) letter was mailed on July 19, 2007, to the Customers that received UniCel Dxl Access
lmmunoassay System Reaction Vessels lot 1201291 and
1201431. The customers were informed that a small
percentage of the reaction vessels (Lot #s: 1201291and
1201431 ) have been found to be defective and there is a
potential for an increase in signal leading to an erroneous
result. The customers were instructed to discontinue use and
discard these lots. A response form was included.
Root Cause
Process design
Distribution Pattern
Nationwide & Canada