Terminated

UniCel Dxl Access Immunoassay Systems Reaction Vessels for Use with UniCel Dxl Access Immunoassay Systems, Part Number : 386167, Beckman Coulter, Inc, Brea CA 92822

FDA Recall Z-1108-2008 · Posted 2008-03-19

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeMMI
Date Initiated2007-07-19
Date Terminated2012-05-04
Product Quantity264 units in the US; 9 units in Canada

Reason for Recall

Erroneous Results - UniCel Dxl Access Immunoassay Systems Reaction Vessels have been found to be defective and there is a potential for an increase in signal leading to an erroneous result.

Firm Action

A Product Corrective Action (PCA) letter was mailed on July 19, 2007, to the Customers that received UniCel Dxl Access lmmunoassay System Reaction Vessels lot 1201291 and 1201431. The customers were informed that a small percentage of the reaction vessels (Lot #s: 1201291and 1201431 ) have been found to be defective and there is a potential for an increase in signal leading to an erroneous result. The customers were instructed to discontinue use and discard these lots. A response form was included.

Root Cause

Process design

Distribution Pattern

Nationwide & Canada