Terminated

ACM Bio-Prep with Merocel pkg/6, Rx, Sterile, REF 0306-708-000S4, Stryker Instruments, Kalamazoo, MI. The femoral sponge is an absorbent that is used to dry the intramedullary canal prior to the application of bone cement. The suction effectively absorbs blood and fluid from the canal. It may provide swabbing/scrubbing action during suction.

FDA Recall Z-1108-2010 · Posted 2010-03-15

Recall StatusTerminated
Product CodeGDY
Date Initiated2010-01-25
Date Terminated2010-10-25
Product Quantity9,274 all products

Reason for Recall

The firm did not have packaging validation to support the labeled shelf life. There is potential for a breach in sterility specific to the sterile packaged femoral sponge which may result in a non-sterile product.

Firm Action

A "Medical Device Recall Notification" dated January 21, 2010 was issued to customers. The notification described the affected product, issue and actions to take. Consignees were asked to forward the notification to all their affected locations. Customers were instructed to immediately check inventory and quarantine any recalled product found. Consignees were instructed to destroy all product and report how much product was destroyed. For questions regarding this recall, please contact Stryker Instruments by calling 1-800-800-4236, ext. 4354

Root Cause

Other

Distribution Pattern

Worldwide Distribution -- United States, Canada, Mexico, Australia, Brazil, Switzerland, Japan, Netherlands, England, UK, Sweden, Hong Kong, Korea, Malaysia, Singapore, Argentina, Spain, and Italy.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →