Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
FDA Recall Z-1109-2017 · Posted 2017-01-27
Advertisement
Recalling FirmImmuno-Mycologics, Inc
Recall StatusTerminated
Product CodeGMD
Date Initiated2016-12-27
Date Terminated2017-03-08
Product Quantity630 units
Reason for Recall
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
Firm Action
Immuno sent an Urgent: Medical Device Recall Letter dated December 27, 2016, to all affected customers.
All customers were notified via emails to stop using the product on December 27th and December 28th. A recall notification and response form was attached to the email. For further questions, please call (405)364-1058.
Root Cause
Material/Component Contamination
Distribution Pattern
Nationwide throughout the US
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.