Boston Scientific Stonetome Stone Removal Device, 5 mm tip, 20 mm cut wire, proximal balloon, sterile, single use only, material number M00535150; catalog number 3515.

FDA Recall Z-1111-06 · Posted 2006-06-27

Recall StatusTerminated
Product CodeLQR
Date Initiated2006-05-19
Date Terminated2007-06-20

Reason for Recall

Lack of assurance of sterility, as the sterile barrier may fail.

Firm Action

Consignees were notified to cease using, and to return, the products via recall letter dated May 19, 2006. International customers were also notified.

Root Cause

Other

Distribution Pattern

Nationwide, Arab Emirate, Australia, Austria, Belgium, Canada, Denmark, Finland, France, germany, Great Britain, Hong Kong, Israel, Italy, Japan, Libya, Mexico, Netherlands, New Zealand, Portugal, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Turkey, Taiwan, Venezuela and Viet Nam.