SPS-1, Static Preservation Solution [1 or 2 liter bags], Sterile; Organ Recovery Systems, One Pierce Place, Itasca, IL 60143
FDA Recall Z-1111-2017 · Posted None
Advertisement
Recalling FirmOrgan Recovery Systems, Inc.
Recall StatusTerminated
Product CodeKDL
Date Initiated2016-12-14
Date Terminated2019-08-14
Product Quantity4,806 bags (3000 1 Liter bags, 1806 2-Liters bags)
Reason for Recall
Organ preservation fluid recalled due to potential for bacterial contamination.
Firm Action
Firm initiated a voluntary removal of Lot PBR-0060-392 on December 14, 2016. and Lot PBR-0074-330 December 16, 2016. Notification occurred via e-mail to each recipient of the lots. On December 18, 2016, the firm provided subsequent posting notice and FAQs on company website. Follow up telephone conversations were performed post e-mail notifications. Firm worked with FDA to develop and coordinate a unified message that would be noticed on the websites of FDA, UNOS, and CDC. Firm also independently contacted the Association of Organ Procurement Organizations and UNOS to assist with messaging for their constituents. In the recall notification emails, customers were instructed to return affected product to Organ Recovery Systems. All affected product will be destroyed after completion of the firm's internal investigation.
Upon learning of the additional lots of potentially contaminated product (PBR-0074-337 and PBR-0060-386), on 1/13/2017 ORS notified their customers via email and phone and updated the posting on their website. The instructions were only to quarantine the product, however. On January 23, 2017, firm began instructing their customers (via email) to return the additional affected lots to ORS. An update to their website was posted January 24, 2017. instructing affected customers to return the lots of product.
For further questions regarding this recall, please call (847) 824-2421.
Root Cause
Material/Component Contamination
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.