Olympus Uretero-reno videoscope, Model No. URF-V2
FDA Recall Z-1116-2022 · Posted None
Advertisement
Recalling FirmOlympus Corporation of the Americas
Recall StatusOpen, Classified
Product CodeNWB
Date Initiated2022-03-25
Product Quantity1
Reason for Recall
The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
Firm Action
On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.
Root Cause
Process change control
Distribution Pattern
US Nationwide distribution in the state of AZ.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.