Rusch Greenlite Single-Use Metal Laryngoscope Blade for Fiber Optic System

FDA Recall Z-1117-2020 · Posted 2020-02-07

Recalling FirmTELEFLEX MEDICAL INC
Recall StatusTerminated
Product CodeCCW
Date Initiated2020-01-09
Date Terminated2023-09-11
Product Quantity338,928 (199,018 US and 139,910 OUS) for all products

Reason for Recall

The products are being recalled due to reports of breakage broke at the point of a welded joint either prior to use during setup or during use inside the patient's mouth.

Firm Action

Letters sent to customers/users to immediately discontinue use and quarantine any products with the affected product code and lot numbers and return them to the manufacturer.

Root Cause

Process control

Distribution Pattern

Distributed nationwide in the US and to Puerto Rico; OUS to more than 20 countries; To the following government VA medical Centers: VAMC - KANSAS CITY MO 4801 E Linwood Blvd Kansas City MO 64128-2218