Terminated

Fisher Healthcare Sure-Vue Serum/Urine hCG-STAT; in vitro diagnostic test; Catalog Number: FHC-A202; Product Codes: FHC-A202-KFI3O & FHC-A202-KFI5O; Inverness Medical Professional Diagnostics, San Diego, CA 92121.

FDA Recall Z-1118-2008 · Posted 2008-05-06

Recalling FirmInnovacon Inc
Recall StatusTerminated
Product CodeJHI
Date Initiated2007-10-01
Date Terminated2011-08-12
Product Quantity1,825,480 units

Reason for Recall

Incorrect results: The Sure-Vue Serum/Urine hCG-STAT test, when interpreted at extended read times, may exhibit sensitivity to patient samples containing hCG at levels well below the cut-off, potentially resulting in a false positive interpretation by the user.

Firm Action

Inverness Medical sent an Urgent Medical Device Alert letter, dated October 1, 2007, to all Innovacon customers for the product via US Postal Service return receipt requested. The field correction notice removed the extended read time of 10 minutes. Customers were instructed to note that although the package insert currently allows interpretation up to 10 minutes, do not read at times greater than 3 minutes for urine and 5 minutes for serum.

Root Cause

Labeling design

Distribution Pattern

Nationwide