Surgical convenience kits labeled as: a) MEDLINE GENERAL UTILITY, Reorder Number DYNDA2524 b) CENTURION MEDICAL PRODUCTS DIALYSIS ON/OFF BUNDLE, Reorder Number DT22705A c) CENTURION MEDICAL PRODUCTS ULTRASOUND PIV BUNDLE, Reorder Number IV8635D d) CENTURION MEDICAL PRODUCTS CUTDOWN TRAY, Reorder Number MNS12595
FDA Recall Z-1118-2023 · Posted 2023-02-10
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeLRO
Date Initiated2022-12-07
Product Quantity630 units
Reason for Recall
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
Firm Action
The firm sent a correction notice to its direct consignees on 12/07/2022 by US mail and email. The notice explained the issue and requested the following:
"Please check you stock for affected kits and over-labeling will be sent to place on kits to ensure that all labeling is accurate. No kits are to be returned back to Medline. All components are not affected. Our records indicate that these were used as a component in the list of kit(s) purchased from Medline Industries, LP."
"If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP."
"If you have any questions, please contact the Recall Department at 866-359-1704."
Root Cause
Under Investigation by firm
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.