Artis One; The Artis One is an angiography system developed for diagnostic interventional procedures including, but not limited to pediatric and obese patients. Procedures that can be performed with the Artis One include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for patient extremities. Additionally, angiographic procedures can be performed in the operating room, image guided surgery by x-ray, by image fusion, and by navigation systems.

FDA Recall Z-1119-2015 · Posted 2015-02-17

Recall StatusTerminated
Product CodeOWB
Date Initiated2015-01-16
Date Terminated2015-04-21
Product Quantity1

Reason for Recall

The possibility exists that the monitor may fail and requires a power circle (shutdown and then power on) to resume operation. The problem is not systematic; but sporadic on single units.

Firm Action

Siemens notified their customer via Customer Safety Advisory Letter AX049/14/S, dated January 16, 2015, and has initiated a fix with Update Instruction AX048/14/S which will be completed by a Siemens service technician. Customers with questions should contact the Customer Care Center at 1-800-888-7436. For questions regarding this recall call 610-219-6300.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution to MI only