Accu-Chek Ultraflex-1 infusion set; 10 mm 31'' (80 cm); Catalog number INF 04540883001 (Distributed U.S. and Canada) and Ref. No 04626524001 (Distributed outside of the the U.S. only).

FDA Recall Z-1120-06 · Posted 2006-07-08

Recall StatusTerminated
Product CodeLZG
Date Initiated2006-03-31
Date Terminated2007-11-06
Product Quantity390,870 in US and Canada plus 10,510 outside N. America

Reason for Recall

The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia

Firm Action

"U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.

Root Cause

Other

Distribution Pattern

Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.