Outer Packaging Label: Protekt Sharps Safety Knives - REF: PG 44620MU, Sterile R, Multi-Use 2.2mm Double Bevel Slit, Angled 6 units, Individual Packaging Label: ProTekt Sharps Safety Knives - REF: PG 44620MU,2.2mm double Bevel Slit, Sterile R, Angled 1 unit, Handle Labeling; 2.5mm slit double bevel

FDA Recall Z-1120-2022 · Posted None

Recall StatusCompleted
Product CodeHNN
Date Initiated2022-02-24
Product Quantity90 units

Reason for Recall

Surgical knives may contain incorrect description printed on handles. Specifically, some surgical knife handles may contain a size of "2.5mm" when the actual size is "2.2mm".

Firm Action

Between the dates of February 16, 2022 - March 10, 2022, Diamatrix Ltd. issued a "Urgent: Medical Device Recall Notification. In addition to notifying consignees about the recalled product, the firm asked consignees to take the following actions: " Discontinue or do not use Diamatrix Item # PG 44620MU ProTekt 2.2mm Double Bevel Slit knives from Lot B923LUC. " Return the product to Diamatrix immediately. Please contact our customer service, 800-867-8081, to obtain a shipping label from Diamatrix. " Acknowledge this Voluntary Recall by returning the included Medical Device Recall Return Response letter within 3 business days of receiving this Recall letter. " A complimentary phone call to the phone numbers mentioned below would assist us with immediate tracking and reports to the FDA.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of MN, TX and the country of Germany.