Open, Classified

(1) BD Pyxis MedStation ES, Catalog No. 323; (2) BD Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) BD Pyxis MedStation ES Tower, Catalog No. 352; (4) BD Pyxis MedStation 4000 Console, Catalog No. 309; (5) BD Pyxis MedStation 4000, Catalog No. 303; (6) BD Pyxis Anesthesia System 4000, Catalog No. 338

FDA Recall Z-1120-2025 · Posted 2025-02-07

Recalling FirmCareFusion 303, Inc.
Recall StatusOpen, Classified
Product CodeBRY
Date Initiated2025-01-08
Product Quantity193067

Reason for Recall

Labeling is insufficient regarding delays in access to medication. The firm is strengthening the label as an additional risk mitigation.

Firm Action

On January 8, 2025, the firm notified customers through an "Urgent Medical Device Correction" letter sent via email and mailed notices. In addition to explaining software issues affecting the device (covered in RES 96055), the letter also provided customers with a list of mitigations that address device downtime as a result of power failure, network failure, or hardware failure. The firm will be updating product labeling and product guides to further strengthen recommendations. On September 25, 2025, additional correction notices were distributed identifying additional affected devices. If you require further assistance contact the firm: Technical Support for technical questions: bd.com/self-service, 1-800-727-6102, Hours: 24 hours, 7 days a week; BD Recall Operations to return the customer response form: [email protected], Fax: (312) 949-0229

Root Cause

Labeling design

Distribution Pattern

Domestic distribution nationwide US. International distribution to North America, Latin America, EMEA, and Greater Asia.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑