Gambro Cartridge Blood Transport System for Hemodialysis Blood Set, product code 109536 - Product Usage: intended for single use in a hemodialysis treatment using the Phoenix Dialysis Delivery Systems.

FDA Recall Z-1121-2021 · Posted None

Recall StatusOpen, Classified
Product CodeFJK
Date Initiated2020-12-30
Product Quantity211155 devices

Reason for Recall

complaints of kinked access lines observed during treatment using cartridge sets for hemodialysis treatments were received.

Firm Action

The firm issued a Safety Alert on 12/30/2020 by mail. The letter explained the problem with photographs, and the hazard involved with using a device with a kinked line. Customers identifying set with kinked tubing may contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Distributor were asked to notify their customers by distributing the letter.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution.