MicoVention Terumo, LVIS Intraluminal Support Device, REF: 213522-CAS, od, 3.5 mm, TL 22 mm, WL 18 mm, Sterile R, CE0297. The intended purpose of the LVIS Device is to serve as a stent-assisted coiling (SAC) device providing support to the embolic coils used in the treatment of wide-necked aneurysms.
FDA Recall Z-1121-2025 · Posted 2025-02-07
Advertisement
Recalling FirmMICROVENTION INC.
Recall StatusOpen, Classified
Product CodeQCA
Date Initiated2024-12-11
Product Quantity11 units
Reason for Recall
Potential for Polytetrafluoethylene (PTFE) material encasing the stent loop may prevent the stent from being pushed out of the introducer.
Firm Action
On December 11, 2024, MicroVention Europe SARL aka Terumo Neuro issued a "Urgent Medical Device Recall" Notification to affected consignees via mail. Terumo Neuro asked consignees to take the following actions:
1. Review Inventory; 2. Inform and Forward Recall Notice; 3. Return Product; 4. Please complete and return the "Customer Acknowledgment Form"; 4. Contact Quality Department ; 5. Replacement or Credit of Returned Devices
Root Cause
Nonconforming Material/Component
Distribution Pattern
US: None OUS: China
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.