MEVION S250-Proton Radiation Therapy Product Usage: Proton Radiation Therapy
FDA Recall Z-1122-2017 · Posted 2017-02-02
Advertisement
Recalling FirmMevion Medical Systems, Inc.
Recall StatusTerminated
Product CodeLHN
Date Initiated2016-12-16
Date Terminated2017-04-13
Product Quantity2 units
Reason for Recall
An error can occur causing Delta corrections to be lost when one setup field is closed and another is opened
Firm Action
Mevion sent an Important Safety Notice to customers on 12/16/16 by email and Salesforce. The letter identified the affected product, problem and actions to be taken. Mevion provided mitigation and will release and install a software patch that removes the possibility that Deltas can be lost between multiple Setup fields. For further information contact [email protected].
Root Cause
Software design
Distribution Pattern
US Nationwide in the states of OK, NJ