BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010
FDA Recall Z-1122-2019 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeJKA
Date Initiated2019-03-07
Date Terminated2020-05-12
Product Quantity6,509,600
Reason for Recall
The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.
Firm Action
Urgent Medical Device Recall notification letters dated 3/7/19 were distributed to customers.
Root Cause
Process change control
Distribution Pattern
US distribution to the states of FL and NJ.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.