Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Intralase FS Laser System. Laser Keratome

FDA Recall Z-1124-03 · Posted 2003-08-12

Recalling FirmIntralase Corp
Recall StatusTerminated
Product CodeGEX
Date Initiated2003-07-15
Date Terminated2004-12-02
Product Quantity70

Reason for Recall

Gantry moves toward patient when 'Home' button is used after procedure.

Firm Action

Firm sent Safety Alert letters to customers on July 15, 2003 by registered mail instructing not to use the HOME button, including a 'Do Not Use' sticker for the machine and an addendum to the Operations Manual.

Root Cause

Other

Distribution Pattern

Nationwide and Japan.