Intralase FS Laser System. Laser Keratome
FDA Recall Z-1124-03 · Posted 2003-08-12
Advertisement
Recalling FirmIntralase Corp
Recall StatusTerminated
Product CodeGEX
Date Initiated2003-07-15
Date Terminated2004-12-02
Product Quantity70
Reason for Recall
Gantry moves toward patient when 'Home' button is used after procedure.
Firm Action
Firm sent Safety Alert letters to customers on July 15, 2003 by registered mail instructing not to use the HOME button, including a 'Do Not Use' sticker for the machine and an addendum to the Operations Manual.
Root Cause
Other
Distribution Pattern
Nationwide and Japan.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.