Accu-Chek Ultraflex-2 infusion set; 8 mm 43'' (110 cm); Catalog number INF 04540948001.

FDA Recall Z-1124-06 · Posted 2006-07-08

Recall StatusTerminated
Product CodeLZG
Date Initiated2006-03-31
Date Terminated2007-11-06
Product Quantity28,170 in US and Canada

Reason for Recall

The luer tube may break at the connection to the pump and, if this is not noticed by the patient, the patient may receive an under delivery of insulin.

Firm Action

"U.S. and International distributors, patients and their physicians were notified via letters dated 3/31/06. Distributors were instructed to search for all part numbers affected by the recall and to provide all affected customers and sub-distributors with the recall notice. Customers were offered a replacement set or replacement with a different model. A press release was also issued on 4/3/06.

Root Cause

Other

Distribution Pattern

Worldwide including USA, Argentina, Austria, Barbados, Belgium, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Honduras, Hungary, Italy, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovenia, Spain, Sweden, Switzerland and United Kingdom.