Easy Access Software releases 1.1, 5.2, 6.2 & 7.2
FDA Recall Z-1125-03 · Posted 2003-08-12
Advertisement
Recalling FirmPhilips Medical Systems
Recall StatusTerminated
Product CodeLLZ
Date Initiated2003-06-27
Date Terminated2005-01-12
Product Quantity46 devices
Reason for Recall
If DICOM fields for body part, description or accession number are changed, change won't update on the EasyWeb client or hospital systems.
Firm Action
On 6/25/03 the firm issued a letter dated 6/15/03 to the affected customers. The letter advises of the problem and advises the customer that the firm will install corrected software in affected units.
Root Cause
Other
Distribution Pattern
Devices were distributed to 23 hospitals and medical centers throughout the US.