cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems
FDA Recall Z-1125-2019 · Posted None
Advertisement
Recalling FirmRoche Diagnostics Corporation
Recall StatusTerminated
Product CodeJJE
Date Initiated2018-12-14
Date Terminated2020-09-04
Product Quantity137 units
Reason for Recall
COBAS INTEGRA 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests
Due to Carry-over
Firm Action
Roche Diagnostics issued on 12/14/18 "Urgent Medical Device Correction (UMDC) via UPS Ground (receipt required) Consignee Instructions:
Manually add the EWCs listed in the Urgent Medical Device Correction (UMDC) on your COBAS INTEGRA 400 plus and cobas c 111 analyzers.
If customers run the test combinations requiring needle assignment (See Table on page 3 of UMDC TP-00508), run those in batch mode until the updated TAS has been loaded on the COBAS INTEGRA 400 Plus analyzer.
Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution.