Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4 --- Common name: Medical image storage device. Class I, 510 (k) exempt.
FDA Recall Z-1126-03 · Posted 2003-08-12
Advertisement
Recalling FirmOlympus America Inc.
Recall StatusTerminated
Product CodeLMB
Date Initiated2003-03-05
Date Terminated2005-03-22
Product Quantity856
Reason for Recall
Software defect; potential under certain circumstances to incorrectly identify endoscopy images that have been uploaded & saved to the network system.
Firm Action
Olympus America, Inc. sent "Urgent-Device Correction" letters, dated 3/5/03, to the medical facilities, advising them of the software malfunction. The letter provides instructions to health care practitioners and that a new software release is being developed to prevent a recurrence.
Root Cause
Other
Distribution Pattern
Medical facilities nationwide and Puerto Rico; and 4 foreign accounts in Canada, Ecuador, Chile & Mexico.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.