Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4 --- Common name: Medical image storage device. Class I, 510 (k) exempt.

FDA Recall Z-1126-03 · Posted 2003-08-12

Recalling FirmOlympus America Inc.
Recall StatusTerminated
Product CodeLMB
Date Initiated2003-03-05
Date Terminated2005-03-22
Product Quantity856

Reason for Recall

Software defect; potential under certain circumstances to incorrectly identify endoscopy images that have been uploaded & saved to the network system.

Firm Action

Olympus America, Inc. sent "Urgent-Device Correction" letters, dated 3/5/03, to the medical facilities, advising them of the software malfunction. The letter provides instructions to health care practitioners and that a new software release is being developed to prevent a recurrence.

Root Cause

Other

Distribution Pattern

Medical facilities nationwide and Puerto Rico; and 4 foreign accounts in Canada, Ecuador, Chile & Mexico.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑