Medline Kits, trays, and packs labeled as follows: a) CATARACT PACK, REF VAL011CTASD; b) CATARACT PACK, REF DYNJ68875F; c) CATARACT PACK, REF DYNJ54869F; d) CATARACT PREP PACK, REF DYNJ44917L; e) E-BLEPH CDS, REF CDS983585G; f) EYE CDS, REF CDS983916G; g) EYE KIT, REF DYKM2222; h) EYE PLASTIC PACK, REF DYNJ47859C; i) K PACK, REF LYN030KAGVD; j) OCULOPLASTY PACK, REF DYNJ64572A; k) OL OCULAR, REF DYNJ906183J; l) OL OCULAR, REF DYNJ906183K; m) P M CATARACT, REF DYNJ44917G; n) P M CATARACT, REF DYNJ44917I; o) P PACK, REF LYN030POGVD; p) R VITRECTOMY, REF EYP087W; q) VITRECTOMY PACK, REF DYNJ68884B
FDA Recall Z-1126-2024 · Posted 2024-02-25
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeOJK
Date Initiated2023-11-15
Product Quantity5724 units
Reason for Recall
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
Firm Action
MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers.
For questions, contact the Recall Department at 866-359-1704 or [email protected].
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.