Instrumentation Laboratory ACL TOP Coagulation Analyzer (Base Model only). Blood analyzer. P/N 280000
FDA Recall Z-1127-2007 · Posted 2007-08-07
Advertisement
Recalling FirmInstrumentation Laboratory Co.
Recall StatusTerminated
Product CodeGKP
Date Initiated2007-06-19
Date Terminated2012-05-23
Product Quantity94
Reason for Recall
ACL TOP unit, when in emergency stop may release liquid during recovery sequence and contaminate sample tube.
Firm Action
Instrumentation Laboratory notiifed the distributor Beckman Coulter on 6/19/2007. Beckman Coulter notified their customers and provided a copy with the Instructions for Recovery Procedure on how to avoid contamination from probe leaks.
Root Cause
Other
Distribution Pattern
Nationwide.