Open, Classified

Cardinal Health Presource Kits: 1) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV20110B; 2) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV20110D 3) KIT,CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA103 4) KIT,CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10T 5) KIT,CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10U 6) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10X 7) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10Y 8) SLH AAA PACK, Catalog Number:SCV13AASLA 9) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAF10 10) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH4 11) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH5 12) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH6 13) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH7 14) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH8 15) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH9 16) MICROVASCULAR FREE FLAP, Catalog Number:SCV30VAFHP 17) MICROVASCULAR FREE FLAP, Catalog Number:SCV30VAFHQ 18) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFHR 19) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFHS 20) NEURO EMBOLISM PACK, Catalog Number:SNE11AN102 21) NEURO EMBOLISM PACK, Catalog Number:SNE11AN10E

FDA Recall Z-1129-2025 · Posted 2025-02-10

Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2024-11-08
Product Quantity973,785 total units

Reason for Recall

Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.

Firm Action

An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Root Cause

Material/Component Contamination

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.