Azurion FlexArm, 722079
FDA Recall Z-1130-2019 · Posted None
Advertisement
Recalling FirmPhilips North America, LLC
Recall StatusTerminated
Product CodeJAA
Date Initiated2019-03-12
Date Terminated2020-10-07
Product Quantity2
Reason for Recall
During manufacturing of an Azurion R2.0 system, it was found out that the cable set was not according to specification. This resulted in damage of the 24V power cable of the detector.
Firm Action
On March 12, 2019, the firm distributed Urgent Medical Device Correction letters to affected consignees, alerting customers that the cable set in the C-arc of the Azurion R2.0 was not according to specifications, which resulted in damage of the 24V power cable of the detector.
Customers were asked to place a copy of the field safety notice together with the documentation of the system until the system is inspected and corrected (if applicable) by Philips.
The firm will inspect the affected systems free of charge and correct as necessary.
If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
Root Cause
Component design/selection
Distribution Pattern
Distributed to one account in Florida, and one international account in Germany.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.