Cardinal Health Presource Kits: 1) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF104; 2) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF105 3) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF106 4) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF107 5) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF108 6) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF109 7) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10V 8) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10W 9) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10Y 10) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF10Z 11) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF111 12) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF112 13) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCP2 14) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCP3 15) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCP4 16) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPA 17) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPB 18) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPC 19) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MFCPD 20) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCPE 21) KIT,MAXILLOFACIAL PRESBYTERIAN, Catalog Number:PWV2MFCPF

FDA Recall Z-1130-2025 · Posted 2025-02-10

Recall StatusOpen, Classified
Product CodeOFX
Date Initiated2024-11-08
Product Quantity973,785 total units

Reason for Recall

Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.

Firm Action

An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Root Cause

Material/Component Contamination

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.