Possis AngioJet XVG Thrombectomy Sets, Model/Order # 105042-001, Sterilized with Ethylene Oxide, Possis Medical, Inc., 9055 Evergreen Boulevard NW, Minneapolis, MN 55433-8003. (These sets are used with the new AngioJet Ultra Console)

FDA Recall Z-1131-2007 · Posted 2007-08-10

Recalling FirmPossis Medical, Inc
Recall StatusTerminated
Product CodeDXE
Date Initiated2007-06-19
Date Terminated2008-03-02
Product Quantity194

Reason for Recall

Mislabeling : Possis AngioJet XVG Catheter Thrombectomy Sets display incorrect information on the outer packaging box and also on the inner sterile tray. The labeling incorrectly states that the unit is indicated for use in coronary vessels. These units should indicated use only for peripheral arteries.

Firm Action

A Urgent Medical Device recall letter dated 19 June 2007, was sent to consignees advising that all unused product be removed and returned to Possis. An acknowledgement of the disposiiton of the product will be requested from consignees.

Root Cause

Other

Distribution Pattern

AR, AZ, CA, FL, GA, IL, IN, KS, KY, MA, ME, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, WA, WI, WV