Recall StatusTerminated
Date Initiated2021-01-15
Date Terminated2023-03-09
Product Quantity49
Reason for Recall
Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
Firm Action
On 01/19/2019, Zimmer Biomet issued Urgent Medical Device Recall notices to distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory will be notified via FedEx. Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
Root Cause
Labeling Change Control
Distribution Pattern
US Nationwide distribution including in the states of MS, MN, KY, FL, MI, OH, AZ, IL, WI, CA, UT, TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.