Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in protective wrap, then into a sterile pouch, 1 device per carton, 1 carton per box.

FDA Recall Z-1131-2022 · Posted None

Recall StatusOpen, Classified
Product CodeOCL
Date Initiated2022-04-20
Product Quantity1236 units

Reason for Recall

Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s.

Firm Action

A recall notification was sent to consignees on 04/20/2022 via the United Parcel Service. The letter instructs the consignees to review the updated guidance on device usage contained in the letter, share the notice to those that need to be aware, and complete the enclosed Customer Confirmation Form and email to [email protected].

Root Cause

Labeling design

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of CA, ND, MA, ID, WI, KS, UT, MN, NY, NJ, GA, MO and the countries of Australia, Belgium, China, Czech Republic, France, Germany, Greece, Israel, Italy, Kazakhstan, Latvia, Netherlands, Poland, Russian Federation, Singapore, Switzerland.