CoCr Femoral Head - A modular head component is used in conjunction with a femoral stem component for replacement of the proximal femur in total hip arthroplasty.

FDA Recall Z-1132-2015 · Posted 2015-02-20

Recalling FirmZimmer Gmbh
Recall StatusTerminated
Product CodeLPH
Date Initiated2014-12-17
Date Terminated2015-09-23
Product Quantity1908

Reason for Recall

Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.

Firm Action

On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.

Root Cause

Package design/selection

Distribution Pattern

Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan