NxStage PureFlow B Solution - Premixed Dialysate for Hemodialysis with NxStage System One

FDA Recall Z-1132-2018 · Posted None

Recalling FirmNxStage Medical, Inc.
Recall StatusTerminated
Product CodeKPO
Date Initiated2017-11-10
Date Terminated2019-12-06
Product Quantity1243400 units

Reason for Recall

Using different fluid formulations of NxStage PureFlow B Solution at the same time during treatment may cause patient health risk. The NxStage PureFlow B Solution is for use with systems that use one premixed dialysate during hemodialysis. The original labeling did not warn against mixing dialysate formulations and did not instruct staff to verify that the fluid is flowing during use.

Firm Action

Customers are advised to : 1. Always make sure when using PureFlow B Solution that fluid is flowing freely from all bags. 2. Recognize that any practice of hanging more than one fluid type at the same time is not an intended use of the NxStage System One" (as it does not monitor fluid composition) and is not a labeled indication of PureFlow B solutions. 3. Please make sure that all users of PureFlow B Solution in your organization and other concerned persons are informed regarding this Field Safety Notice. 4. Complete the enclosed customer reply form and return it to NxStage. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices.

Root Cause

Labeling design

Distribution Pattern

US nationwide distribution