FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics, REF: RFIT-ASY-0126, and RFIT-ASY-0127, when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/410851; BACT/ALERT FN Plus/410852; BACT/ALERT PF Plus/410853

FDA Recall Z-1133-2019 · Posted None

Recall StatusTerminated
Product CodePEN
Date Initiated2019-03-07
Date Terminated2022-08-24
Product Quantity20350

Reason for Recall

Increased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles.

Firm Action

On 03/07/19, Urgent: Medical Device Recall (Correction) notices were mailed to customers. Customers were informed that the false positive results are primarily Enterobacteriaceae results with a lower rate of concurrent E. coli results. If the affected BCID panel is used with BACT/ALERT blood culture bottles from the lots listed, then positive results for Escherichia coli and Enterobacteriaceae (if occurring without a specific genus/species reported other than E. coli; i.e. negative for all of the following: Enterobacter cloacae complex, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens) should be confirmed by another method prior to reporting the test results. Confirmation is not required if the Enterobacteriaceae detected result is accompanied by a detected result for Enterobacter cloacae complex, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, or Serratia marcescens. Customers were asked to complete and return Acknowledgement of Receipt forms.

Root Cause

Material/Component Contamination

Distribution Pattern

MA, NH, VT, CT, NJ, NY, PA, DE, DC, VA, MD, WV, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, ND, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, WY, ID, UT, AZ, NM, NV, CA, HI, OR, WA, AK, BM

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.