Terminated

SenTec Digital Monitor, Model SDM, is a portable stand-alone patient monitor indicated for continuous, noninvasive patient monitoring of carbon dioxide partial pressure (PCo2), oxygen partial pressure (POs), functional oxygen saturation (PsO2), and pulse rate (PR), using either a single digital sensor (V-Sign Sensor 2 or OxiVenT sensor).

FDA Recall Z-1134-2021 · Posted None

Recalling FirmSenTec AG
Recall StatusTerminated
Product CodeLKD
Date Initiated2020-12-10
Date Terminated2026-08-31
Product Quantity15 devices

Reason for Recall

Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards.

Firm Action

The firm initiated their correction by letter to the direct accounts on 12/10/2020. The letter explains the problem and provides additional instruction manuals and a sticker to address the use of the product in the home care environment.

Root Cause

Device Design

Distribution Pattern

IL, PA, FL, NJ, GA, MA