Terminated

VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.

FDA Recall Z-1136-2021 · Posted None

Recall StatusTerminated
Product CodeDIP
Date Initiated2021-01-14
Date Terminated2022-12-27
Product Quantity4,062 cartons of 1,000 tips (Totaling 4,062,000 tips)

Reason for Recall

A deformity affecting a specific lot of VITROS Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

Firm Action

On 01/18/2021, Ortho Clinical Diagnostics issued an Urgent Product Correction Notification via e-mail due to a a deformity in a specific lot of VITROS Chemistry Products VersaTips. The affected product is to be no longer distributed and discarded.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada.