TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
FDA Recall Z-1136-2026 · Posted 2026-01-21
Advertisement
Recalling FirmSysmex America, Inc.
Recall StatusOpen, Classified
Product CodeLXG
Date Initiated2025-12-15
Product Quantity44 units
Reason for Recall
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
Firm Action
A Product Notification Field Action was sent to customers on 12/15/25.
Actions for customers
1. To prevent injury, never place your hands around the lift section at any time.
2. In addition, please follow these instructions when handling rack jams or similar errors:
" Pay special attention not to insert your hands around the lift section during any
intervention.
" If a rack, tube, or any object becomes lodged around the lift section, turn off the power.
After confirming that the lift section has moved to the bottom dead point, carefully remove the rack and blood collection tubes.
" If the rack or blood collection tubes are jammed in the lift section and cannot be removed with one hand, contact Sysmex for assistance.
3. Sysmex will implement a software update to improve lift position control to prevent unintended sudden drops of the lift section by ensuring that, after an error occurs, the lift automatically returns to the home position. Our service team will contact you to schedule the installation of software version 00-18 by 12/31/2025.
4. Software update acknowledgement
a. To assist Sysmex with the field safety communication, please acknowledge the
communication by proceeding to the URL below to complete and submit the
Acknowledgement Form. See Figure 1.
URL: https://pages.sysmex.com/TS-10-Acknowledgement.html
b. On the Acknowledgement Form at the URL, you will be directed to enter First/Last Name, Role/Title, Email Address, Phone #, Facility Name, Street Address, City, State / Province, Zip Code, Device Serial Number and acknowledge receipt of the notice by selecting Yes.
c. Once the Acknowledgement Form is completed, a confirmation email will be sent to your email address.
NOTE: If you do not see the email confirmation, check the SPAM or JUNK folders of your
email box.
5. Please distribute this Product Notification as appropriate to your laboratory staff.
6. File this Product Notification as a part of your laboratory s Quality System as required.
Questions
Root Cause
Software design
Distribution Pattern
US Nationwide.
Official Source