MultiDiagnost Eleva Flat Detector X-Ray System; Part Number: SW 9896 001 43361; Philips Medical Systems, Bothell, WA, 98021
FDA Recall Z-1137-2008 · Posted 2008-05-28
Advertisement
Recalling FirmPhilips Medical Systems North America Co. Phillips
Recall StatusTerminated
Product CodeJAA
Date Initiated2007-11-29
Date Terminated2010-09-17
Product Quantity9 units were distributed to hospitals in the US, 6 were installed, 3 were not installed or in use.
Reason for Recall
Delay in Imaging: MultiDiagnost Eleva with Flat Detector x-ray system cannot perform fluoroscopy procedures while certain image file housekeeping operations are performed. Fluoroscopy cannot be performed and new exposures are not able to be acquired until the housekeeping operations are complete which may take several minutes. The loss of system availability could occur during an interventional
Firm Action
On 11/29/07 Philips Medical Systems sent a letter dated 11/28/07 to customers stating that if users perform certain image file housekeeping operations, fluoroscopy cannot be performed and new exposures cannot be acquired until those operations are complete. It also instructs that the unavailability of fluoroscopy and exposure can be avoided by not pressing the "Delete All Images" button until any pending Autopush operation is complete. FCO 70800067 will be issued to service representatives to install new software.
Root Cause
Software design
Distribution Pattern
USA Distribution - CO, MN, NC, NJ and TX.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.