NovaStar TS, Anti-Asphyxia Valve, Reusable, Material Numbers MP01576 (Size S), MP01577 (Size M), and MP01578 (Size L)
FDA Recall Z-1137-2023 · Posted 2023-02-24
Advertisement
Recalling FirmDraeger Medical, Inc.
Recall StatusOpen, Classified
Product CodeCBK
Date Initiated2023-01-23
Product Quantity16 units
Reason for Recall
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
Firm Action
An Urgent Medical Device Correction notification letter dated January 2023 was sent to customers.
Actions to be taken:
Use of the mask is contraindicated for patients and other persons in close vicinity to patients using the masks, that have implanted devices that may be affected by magnets, including but not limited to:
Pacemakers
Implantable cardioverter defibrillators (ICD)
Neurostimulators
Magnetic metallic implants/electrodes/valves placed in upper limbs, torso, or higher (i.e., neck and
head)
Cerebral spinal fluid (CSF) shunts (e.g., ventriculoperitoneal (VP) shunt)
Aneurysm clips
Embolic coils
Intracranial aneurysm intravascular flow disruption devices
Metallic cranial plates, screws, burr hole covers, and bone substitute devices
Metallic splinters in the eye
Ocular implants (e.g., glaucoma implants, retinal implants)
Certain contact lenses with metal
Implants to restore hearing or balance that have an implanted magnet (such as cochlear implants, implanted bone conduction hearing devices, and auditory brainstem implants)
Magnetic denture attachments
Metallic gastrointestinal clips
Metallic stents (e.g., aneurysm, coronary, tracheobronchial, biliary)
Implantable ports and pumps (e.g., insulin pumps)
Hypoglossal nerve stimulators
Devices labeled as MR (Magnetic Resonance) unsafe
Magnetic metallic implants not labeled for MR or not evaluated for safety in a magnetic field
Apart from the contraindication, you can continue to use NovaStar masks, if you ensure that the magnets are kept at least 15 cm (approx. 6 inches) away from any other medical implants or medical devices that can be impacted by the magnetic fields.
According to our records, your facility has received NovaStar TS masks material number MP01576, MP01577, MP01578, MP01579, MP01580, and/or MP01581, manufactured by Draegerwerk AG & Co. KGaA.
Please add a copy of this Urgent Medical Device Correction Notification to every box of masks you have in your facilit
Root Cause
Error in labeling
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.