Rifton Equipment, Low Base TRAM. Model/Catalog Number: K320. The TRAM is a powered lift device for transferring clients in a seated position or assisting in a sit-to-stand transfer and gait training. The low base variant includes smaller casters and lower base frame so that the legs fit underneath beds.

FDA Recall Z-1138-2026 · Posted 2026-01-22

Recall StatusOpen, Classified
Product CodeFSA
Date Initiated2025-12-18
Product Quantity1,096 units

Reason for Recall

Potential for fraying on the body support strap.

Firm Action

Community Products notified consignees via email on about 12/18/2025. Consignees were instructed to notify all customers, if further distributed, of the recall, inspect all affected devices for visible fraying, and cease use if any is observed. Consignees were requested to provide contact information for the individual responsible for performing the replacement. If devices were further distributed, consignees were advised to manage all customer outreach, inspection and replacement. The other option was to provide Community Products with customer contact information, and Rifton will manage the process. Lastly, consignees were requested to confirm receipt of the notification.

Root Cause

Process change control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of UAE, Australia, Canada, EU, Israel, Hong Kong, Japan, Korea, New Zealand, Russia, Saudi Arabia, and UK.