stryker Precision Thin, Catalog Number REF 2296-003-155; Micro Oscillating and Sagittal Blades
FDA Recall Z-1139-2025 · Posted 2025-02-11
Advertisement
Recalling FirmStryker Corporation
Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2025-01-08
Product Quantity238 units
Reason for Recall
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Firm Action
Stryker issued an URGENT MEDICAL DEVICE recall notice to its consignees on 01/08/2025 via FedEx. The notice explained the issue, potential risk, and requested the affected items be quarantined pending return. Stryker will issue a shipping label to the consignee once the response form has been received. Distributors were directed to forward the notice to all those to whom the affected devices were distributed.
Root Cause
Process control
Distribution Pattern
Worldwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.