B-K Medical Intraoperative Transducer Type 8814, diagnostic ultrasound transducer, B-K Medical Systems, Inc., Wilmington, MA 01887
FDA Recall Z-1140-2008 · Posted 2008-06-10
Advertisement
Recalling FirmB & K Medical Systems, Inc.
Recall StatusTerminated
Product CodeIYN
Date Initiated2008-01-24
Date Terminated2008-06-23
Product Quantity7
Reason for Recall
Crack: Incomplete glue joint in the device housing may affect electrical safety or sterilization.
Firm Action
B-K Medical issued letter notification on 1/23/08 to accounts via Fed Ex, advising them not to use the unit until a representative can inspect and exchange, if unit is found defective. Instructions are also provided to continue use of the device with a sterile cover until site visit.
Root Cause
Process control
Distribution Pattern
USA Distribution: MA, MI, PA, TX, and OR.