Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.
FDA Recall Z-1140-2019 · Posted None
Advertisement
Recalling FirmPhilips Medical Systems Nederlands
Recall StatusTerminated
Product CodeIZI
Date Initiated2019-02-19
Date Terminated2020-10-14
Product Quantity7209 total
Reason for Recall
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
Firm Action
The firm previously notified customers on September 24, 2014 as part of Recall Z-1280-2016. On February 21, 2019, the firm sent an updated recall letter to newly identified affected consignees. Customers were alerted that bent pedal of footswitch might cause interruption of live Flouroscopy imaging or exposure.
Action to be taken by Customer/User:
1. Before starting a procedure, the user should check the footswitch for possible bent pedals. If bent pedals are found the procedure should not continue and Local Service should be contacted.
2. Do not use the footswitch on an anti-fatigue mat or in the pedestal.
3. Please ensure that all staff with access to the affected systems are informed
of the content of this Medical Device Correction.
4. Include this Medical Device Correction together with the documentation of the
system until the system has been corrected by Philips.
The firm stated that the problem will be resolved by an update of the mechanical hardware of the footswitch which will prevent the pedals of the footswitch from becoming bent. The mechanical hardware was available as of February 2019. Customers will be notified by their local Philips representative to schedule a date for installation of the mechanical hardware update.
Technical Support Line: 1-800-722-9377
Root Cause
Component design/selection
Distribution Pattern
US Nationwide Distribution in the states of CA, FL, MI, OR, PA, TX and WA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K050151MULTIDIAGNOST ELEVAPhilips Medical Systems North America, Inc. · 2005-02-09K041949ALLURA XPER FD10Philips Medical Systems North America Co. · 2004-07-30K033737ALLURA XPER FD20Philips Medical Systems North America, Inc. · 2003-12-09K031333PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.2Philips Medical Systems North America, Inc. · 2003-05-13K984545RELEASE 2 OF PHILIPS INTEGRIS SERIES (V5000, BV5000, H5000, BH5000, HM2000) SYSTEMSPhilips Medical Systems North America, Inc. · 1999-02-25
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.