Embrace Enteral Pump
FDA Recall Z-1141-2007 · Posted 2007-08-08
Advertisement
Recalling FirmRoss Products Division Abbott Laboratories
Recall StatusTerminated
Product CodeLZH
Date Initiated2007-06-21
Date Terminated2012-09-20
Product Quantity9,600 units
Reason for Recall
Continued use of the pump may cause the door pivot point on the back of the pump to crack or break. This will cause the pump to shut down and stop the initiation of enteral feeding.
Firm Action
Consignees were notified by letter dated 6/25/2007. The firm sent a revised letter on 07/26/2007.
Root Cause
Equipment maintenance
Distribution Pattern
The product was distributed throughout the US.