The Companion ClearStar Pump System, for enteral feeding.
FDA Recall Z-1142-2007 · Posted 2007-08-07
Advertisement
Recalling FirmRoss Products Division Abbott Laboratories
Recall StatusTerminated
Product CodeLZH
Date Initiated2007-06-21
Date Terminated2012-09-26
Product Quantity27,814
Reason for Recall
The transducer adapter bracket on the pump could crack or break and potentially result in under delivery of the enteral product.
Firm Action
Consignees were notified by letter dated 6/25/2007. A revised letter was sent 07/26/2007.
Root Cause
Equipment maintenance
Distribution Pattern
Worldwide, including USA and Canada.