The Companion ClearStar Pump System, for enteral feeding.

FDA Recall Z-1142-2007 · Posted 2007-08-07

Recall StatusTerminated
Product CodeLZH
Date Initiated2007-06-21
Date Terminated2012-09-26
Product Quantity27,814

Reason for Recall

The transducer adapter bracket on the pump could crack or break and potentially result in under delivery of the enteral product.

Firm Action

Consignees were notified by letter dated 6/25/2007. A revised letter was sent 07/26/2007.

Root Cause

Equipment maintenance

Distribution Pattern

Worldwide, including USA and Canada.