Recall StatusTerminated
Date Initiated2011-10-21
Date Terminated2014-02-04
Product Quantity48
Reason for Recall
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the
Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co
Firm Action
A field action was initiated on October 21, 2011, and completed March 9, 2012, . The issue was corrected to all affected systems by applying the missing label onto the gantry. No recall letter was issued.
For further questions please call (262) 513-4122.
Root Cause
Labeling design
Distribution Pattern
Worldwide Distribution - Nationwide Distribution including the states of FL, IL, IN, LA, MI, MN, NJ, NM, NY, OH, PR, SC, TX and VA and the countries of Vietnam, Latvia, Italy, Israel, Indonesia, Hong Kong, United Kingdom, Finland, Spain, Czech Republic, Canada, Bahrain and Australia.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K052412INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQGE Healthcare · 2005-09-16K052157INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQGe Medical Systems, LLC · 2005-08-25K050489INNOVA 2100Ge Medical Systems, LLC · 2005-03-11K042053INNOVA 4100, INNOVA 3100Ge Medical Systems, Inc. · 2004-08-26K033244DIGITAL FLUOROSCOPIC IMAGING SYSTEM, MODEL INNOVA 4100 WITH TILT TABLE OPTIONGe Medical Systems, Inc. · 2003-10-16
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.