Recall StatusTerminated
Date Initiated2020-01-29
Date Terminated2024-04-16
Product QuantityTotal 791 Traveler (0 US and 791 OUS)
Reason for Recall
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
Firm Action
On 01/29/2020, the firm field personnel and Affiliates directly contacted the first customers beyond the firm's control via phone, email, or personal visit to facilitate rapid notification. The "URGENT MEDICAL DEVICE RECALL" Notification informed customers that specific lots of its Coronary Dilatation Catheters with diameters of 4.0 mm, 4.5 mm, and 5.0 mm may exhibit difficulty or inability to deflate the balloon due to weaker material proximal to the balloon bond resulting from excess heat exposure during manufacturing, and the potential risks with the use of the affected products include air embolism, thrombosis, myocardial infarction and additional intervention.
Healthcare Professionals will be instructed to:
" Reference the attached list of affected part numbers and lot numbers
" Immediately stop using affected devices from these lots
" Review your inventory, complete and return the attached Effectiveness
Check Form
" Return all unused affected product to Abbott
" Share this notification with other relevant personnel in their organization
The action the Recalling Firm is taking:
* Stopped shipping affected lots
* Will implement appropriate corrective actions to ensure product performance
* Field Representatives can assist in identifying and returning affected devices
* Will work with customers to replace returned units with similar devices, pending availability.
For any questions, contact Customer Service Department at (800) 227-9902.
Root Cause
Process change control
Distribution Pattern
Worldwide Distribution: US (Nation-wide) and OUS (countries of):
Algeria Jordan
Argentina Korea
Armenia Kuwait
Aruba Lebanon
Australia Malaysia
Austria, Mauritius
Bahrain Mexico
Bangladesh Morocco
Belgium Netherlands
Brazil New Zealand
Brunei Norway
Bulgaria Oman
Canada Pakistan
Chile Panama
China Peru
Colombia Poland
Costa Rica Portugal
Croatia R¿union
Cyprus Romania
Czech Republic Russian Federation
Denmark Saudi Arabia
Dominican Republic Singapore
Egypt Slovakia
Estonia Slovenia
Finland South Africa
France Spain
Germany Sri Lanka
Greece State of Palestine
Hong Kong Sweden
Hungary Switzerland
Iceland Taiwan, Province of China
India Thailand
Indonesia Tunisia
Iran Turkey
Iraq Ukraine
Ireland United Arab Emirates
Israel United Kingdom
Italy Uruguay
Japan Viet Nam
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.