IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549; dental cement

FDA Recall Z-1142-2024 · Posted 2024-02-16

Recalling FirmVoco GmbH
Recall StatusOpen, Classified
Product CodeEMA
Date Initiated2024-01-25
Product Quantity1530 units

Reason for Recall

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.

Firm Action

Voco issued a field safety notice to its consignees on 01/25/2024 via email. The notice explained the problem, risk, and requested the product be returned. Those who have sold the product were directed to notify their customers and collect the affected product. For questions contact Mr. Michael Cyll at [email protected].

Root Cause

Process control

Distribution Pattern

Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.