Terminated

Kimberly-Clark Microcuff Endotracheal Tube, Pediatric Oral/Nasal Magill, sizes 3.0mm, 3.5mm, 4.0mm, and 4.5mm, Rx Only, Sterile, Manufacturer: Kimberly-Clark NV Belgicastraat 13, 1930 Zaventem, Belgium, Made in Malaysia

FDA Recall Z-1143-2007 · Posted 2007-08-08

Recall StatusTerminated
Product CodeBTR
Date Initiated2007-06-04
Date Terminated2008-06-12
Product Quantity4376 cases (43760 units/total for both products)

Reason for Recall

The tubes may kink during use which may impede or prevent ventilation.

Firm Action

Consignees were notified by letter sent via overnight mail on/about June 4, 2007. A follow-up letter was sent to consignees via fax on June 5, 2007. Consignees were instructed to cease distributing the product and return all affected product in inventory to Kimberly-Clark. Distributors were instructed to notify their accounts of the recall. A Product Recall Response Sheet (Packing Slip) was attached to the recall notice.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Canada, Europe, Australia, Chile, Hong Kong, Singapore, Thailand, India, Malaysia and Saudi Arabia.