Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories
FDA Recall Z-1144-2019 · Posted None
Advertisement
Recalling FirmPhilips Medical Systems (Cleveland) Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2018-07-03
Date Terminated2020-04-17
Product Quantity2
Reason for Recall
Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
Firm Action
On June 29. 2018, the firm sent Customer Information Letters to affected customers. The letter informed customers that Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images.
Customers were advised to review the user documentation provided with your system related to Quality Assurance checks. If artifacts are detected during review of the images produced by these checks, and are not removed with the recommended steps identified within your user documentation, please contact your service representative before resuming clinical use.
Root Cause
Device Design
Distribution Pattern
Distributed only to China and UK.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.